X-LCâ„¢ 3075UV/3070UV
To ensure the most sensitive and stable X-LCtm detection these models utilize a Czerny-Turner monochromator covering a wide wavelength range from 190 to 900 nm with deuterium and halogen lamps (X-LCtm 3070UV) and from 190 to 600 nm with a
single deuterium lamp (X-LCtm 3075UV). The excellent optical characteristics and program capabilites of the X-LCtm 3070/3075UV are combined in one compact package, making the X-LCtm 3070/3075UV the optimal tool for X-LCtm UV/Vis detection. To efficiently detect the much narrower peaks that are obtained using X-LCtm, the
X-LCtm 3070/3075UV is capable of data acquisitions at frequencies up to 100 data points per second.
Xtreme Sensitivity
An extrenely low dispersion cell and a fast response enable detection of sharp and narrow peaks eluted a few seconds with excellent accuracy and reproducibility.
The X-LCtm 3070/3075UV detectors extremely fast time constant capability provides precise peak shapes with minimized noise. Three modes, FAST, STD and SLOW are available to match peak width. The graph shows that very narrow peaks can be precisely recorded using the FAST mode, while peaks recorded using a slow time constant (as typical in conventional HPLC) show tailing and are right-shifted.