Micro-LC system: Solvent delivery

Micro21 LC-01/02: Capillary Liquid Chromatograph

Seamless coupling with MS

Precise flow control at µL/min-order flow rate range makes possible the introduction of the entire amount of eluate from the column into mass spectrometry detectors without splitting.

Gradient starting from 1 µL/min

Pulse-free high precision flow control enables a high-pressure mixing gradient from 1 µL/min by using two built-in pumps. The mixer section is a dynamic mixing system, and its chamber capacity is 20 µL. The degasser, using  the new material Teflon AF, achieves high degassing efficiency even at a low degassing volume of 145µL, thereby shortening the time required for solvent exchange.

Gradient starting from 1 µL/min:

 

 

Gradient starting from 1 µL/min:

 

 

Micro21 PU-01: Single Solvent Delivery Pump for Capillary LC

Excellent pump performance

A 2µL pump head, 1 mm external diameter plunger, and a pulse correction mechanism ensure the most reliable, accurate and stable solvent delivery at low µL/min flow rates.  This results in a retention time and area value precision on par with conventional-scale HPLC.  Recommended flow rate ranges are 0.2 to 200 µl/min for isocratic and 1.0 to 200 µl/min for gradient